Engineering Design of a Special Purpose Functional
Magnetic Resonance Scanner Magnet
Alice Borceto1 ,Daniele Damiani2, Andrea Viale1,
Franco Bertora1, Member, IEEE, and Roberto Marabotto2
1Robotics, Brainand Cognitive Sciences
Dept. of IstitutoItalianodi Tecnologia, 16163 Genova, Italy
([email protected])
2ASGSuperconductors S.p.A., 16152Genova, Italy
([email protected])
Abstract - A 2 T open MRI scanner for fMRI investigations of subjects maintaining a natural stance and free access to the environment is presented. The self-shielded multi-coil magnet is composed from solenoids optimized in positions and cross-sectional shape. It provides a room temperature gap of 800 mm at the subject shoulders, tapering down to 600 mm at its narrowest point. The system consists of symmetrical magnetic poles, five NbTi superconducting coils each, maintained at 4.2 K in a liquid helium cryostat with a re-condenser. Windings, coil formers, mechanical structure, cryostat, vacuum chamber, thermal shield, mechanical supports, and cryogenics have been modeled in detail using Catia CAD software, accounting for the interactions among components and the constraints posed by winding and assembly methods. The cable characteristics have been chosen based on simulations of thermodynamic stability in operating conditions; self-protection against quench has been provided by inductance subdivision with bypass diodes
Keywords - Field homogeneity, magnetic resonance imaging, magnets, MRI
IEEE/CSC & ESAS European Superconductivity News Forum (ESNF) No. 22 October/November 2012. Reference No.ST315; Categories 4, 6.
The published version of this preprint appeared in IEEE Transactions on Applied Superconductivity 23, 4400205 (June 2013).